Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:COUNt?
This command queries the number of peaks that have been found during a peak
search.
The actual number of peaks that have been found may differ from the number of peaks
you have set to be found because of the peak excursion.
Suffix:
<n>
<m>
Return values:
<NumberOfPeaks>
Example:
Usage:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks[:IMMediate] <Peaks>
This command initiates a peak search.
Suffix:
<n>
<m>
Parameters:
<Peaks>
Example:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE <MaxNoPeaks>
This command defines the maximum number of peaks that the R&S FPL1000 looks for
during a peak search.
Suffix:
<n>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
See
"Display Marker Numbers"
irrelevant
irrelevant
CALC:MARK:FUNC:FPE:COUN?
Queries the number of peaks.
Query only
Window
Marker
This parameter defines the number of peaks to find during the
search.
Note that the actual number of peaks found during the search
also depends on the peak excursion you have set with
CALCulate<n>:MARKer<m>:PEXCursion.
Range:
1 to 200
CALC:MARK:PEXC 5
Defines a peak excursion of 5 dB, i.e. peaks must be at least 5
dB apart to be detected as a peak.
CALC:MARK:FUNC:FPE 10
Initiates a search for 10 peaks on the current trace.