10.6.11.5
Example: measuring the harmonic distortion
//-------------Configuring the Measurement-------------
*RST
//Resets the instrument.
INIT:CONT OFF
//Turns on single sweep mode.
CALC:MARK:FUNC:HARM ON
//Turns on the harmonic distortion measurement.
CALC:MARK:FUNC:HARM:NHAR 3
//Defines three harmonics to be found.
CALC:MARK:FUNC:HARM:BAND:AUTO OFF
//Turns off automatic bandwidth selection.
CALC:MARK:FUNC:HARM:PRES
//Determines the ideal configuration.
//-------------Performing the Measurement-------------------
INIT;*WAI
//Initiates the measurement and finishes the sweep.
//-------------Retrieving the Results------------------------
CALC:MARK:FUNC:HARM:LIST?
//Queries the position of the harmonics.
CALC:MARK:FUNC:HARM:DIST? TOT
//Queries the total harmonic distortion.
10.6.12
Measuring the third order intercept point
Determining the TOI
Programming example: measuring the TOI
10.6.12.1
Determining the TOI
All remote control commands specific to TOI measurements are described here.
Useful commands for TOI measurements described elsewhere
CALCulate<n>:DELTamarker<m>:X
CALCulate<n>:DELTamarker<m>:X:RELative?
CALCulate<n>:DELTamarker<m>:Y?
CALCulate<n>:MARKer<m>:X
CALCulate<n>:MARKer<m>:Y?
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
....................................................................................................... 717
on page 841
on page 858
on page 844
on page 858
Remote commands
............................................................... 720
on page 857