DIAGnostic:SERVice:INPut:PULSed:CFRequency <Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect]
Manual operation:
DIAGnostic:SERVice:INPut:PULSed:MCFRequency <Frequency>
This command sets the calibration frequency for frequencies greater than 7 GHz. This
command only takes effect if a microwave calibration signal is selected for input
(DIAGnostic:SERVice:INPut[:SELect]
Parameters:
<Frequency>
Example:
DIAGnostic:SERVice:INPut:RF[:SPECtrum] <Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth>
DIAGnostic:SERVice:INPut[:SELect] <Signal>
This command activates or deactivates the use of an internal calibration signal as input
for the R&S FPL1000.
Parameters:
<Signal>
Example:
Manuel d'utilisation 1179.5860.09 - 12
See
"Calibration Frequency RF"
*RST:
7 GHz
Default unit: Hz
DIAG:SERV:INP:PULS:MCFR 7,1 GHz
NARRowband | BROadband
NARRowband
Narrowband signal for power calibration of the frontend.
BROadband
Broadband signal for calibration of the IF filter.
CALibration
Uses the calibration signal as RF input.
MCALibration
Uses the calibration signal for the microwave range as RF input.
RF
Uses the signal from the RF input.
*RST:
RF
DIAG:SERV:INP CAL
Uses the calibration signal as RF input.
Remote commands
Configuring the R&S FPL1000
on page 961.
on page 533
on page 961).