R&S
FPL1000
Example:
Manual operation:
[SENSe:]CORRection[:STATe] <State>
This command turns correction of measurement results (normalization) on and off.
The command is available after you have created a reference trace for the selected
measurement type with
on page 799.
This command is only available if the Tracking Generator is active (see
OUTPut<up>[:STATe]
Parameters:
<State>
Example:
Manual operation:
[SENSe:]CORRection:COLLect[:ACQuire] <MeasType>
This command initiates a reference measurement (calibration). The reference mea-
surement is the basis for the measurement normalization. The result depends on
whether a reflection measurement or transmission measurement is performed (see
[SENSe:]CORRection:METHod
To obtain a correct reference measurement, a complete sweep with synchronization to
the end of the sweep must have been carried out. This is only possible in the single
sweep mode.
This command is only available if the Tracking Generator is active (see
OUTPut<up>[:STATe]
Setting parameters:
<MeasType>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
SOUR:INT:FREQ:OFFS 10kHz
See
"Frequency Offset"
[SENSe:]CORRection:COLLect[:ACQuire]
on page 796).
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
1
CORR ON
Activates normalization.
See
" Normalization state"
on page 800).
on page 796).
THRough | OPEN
THRough
"TRANsmission" mode: calibration with direct connection
between generator and device input
"REFLection" mode: calibration with short circuit at the input
OPEN
only allowed in "REFLection" mode: calibration with open input
Remote commands
on page 301
on page 302