INITiate<n>[:IMMediate]
surements",
10.6.11.4
Retrieving results
The following commands retrieve the results of the harmonic distortion measurement.
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?
This command queries the total harmonic distortion of the signal.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<m>
Query parameters:
<Result>
Return values:
<DistortionPct>
<DistortionDb>
Usage:
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST
This command queries the position of the harmonics.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<m>
Return values:
<Harmonics>
Manuel d'utilisation 1179.5860.09 - 12
on page 618
INITiate<n>:CONTinuous
Window
Marker
TOTal
Query only
INITiate<n>:CONTinuous
Window
Marker
Returns one value for every harmonic.
The first value is the absolute power of the first harmonic. The
unit is variable.
The other values are power levels relative to the first harmonic.
The unit for these is dB.