Télécharger Imprimer la page

Rohde & Schwarz FPL1000 Manuel D'utilisation page 889

Masquer les pouces Voir aussi pour FPL1000:

Publicité

Les langues disponibles

Les langues disponibles

®
R&S
FPL1000
Usage:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:QFACtor?
This command queries the Q factor of n dB down measurements.
Suffix:
<n>
<m>
Return values:
<QFactor>
Usage:
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult?
This command queries the distance of the n dB down markers from each other.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<m>
Return values:
<Distance>
Example:
Usage:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
Query only
See
"n dB down Value"
irrelevant
irrelevant
Query only
INITiate<n>:CONTinuous
irrelevant
irrelevant
The result depends on the span.
In case of frequency domain measurements, the command
returns the bandwidth between the two n dB down markers in
Hz.
In case of time domain measurements, the command returns the
pulse width between the two n dB down markers in seconds.
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:NDBD ON
Switches on the n dB down function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:NDBD:RES?
Outputs the measured value.
Query only
See
"n dB down Marker State"
Setting basic spectrum RF measurement parameters
on page 381
on page 620.
on page 381
Remote commands
889