Manual operation:
DIAGnostic:SERVice:STESt:RESult?
This command queries the self-test results.
Return values:
<Results>
Example:
Usage:
SOURce<si>:TEMPerature:FRONtend
This command queries the current frontend temperature of the R&S FPL1000.
During self-alignment, the instrument's (frontend) temperature is also measured (as
soon as the instrument has warmed up completely). This temperature is used as a ref-
erence for a continuous temperature check during operation. If the current temperature
deviates from the stored self-alignment temperature by a certain degree, a warning is
displayed in the status bar indicating the resulting deviation in the measured power lev-
els. A status bit in the STATUs:QUEStionable:TEMPerature register indicates a
possible deviation.
Suffix:
<si>
Return values:
<Temperature>
Example:
10.10.3
Working with transducers
The following commands configure and control transducer factors.
Useful commands for transducer management described elsewhere
MMEMory:SELect[:ITEM]:TRANsducer:ALL
Manuel d'utilisation 1179.5860.09 - 12
See
"NONE"
on page 533
See
"Calibration Frequency RF"
See
"Calibration Frequency MW"
String of data containing the results.
The rows of the self-test result table are separated by commas.
DIAG:SERV:STES:RES?
would return, e.g.
"Total Selftest Status:
PASSED", "Date (dd/mm/yyyy): 09/07/2004 TIME:
16:24:54", "Runtime: 00:06", "...
Query only
irrelevant
Temperature in degrees Celsius.
SOUR:TEMP:FRON?
Queries the temperature of the frontend sensor.
Remote commands
Configuring the R&S FPL1000
on page 533
on page 533
on page 928
968
968