CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y <RefPointLevel>
This command defines the vertical position of the fixed delta marker reference point.
The coordinates of the reference may be anywhere in the diagram.
Suffix:
<n>
<m>
Parameters:
<RefPoint>
Example:
Manual operation:
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y:OFFSet <Offset>
This command defines a level offset for the fixed delta marker reference point.
Suffix:
<n>
<m>
Parameters:
<Offset>
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe] <State>
This command activates or deactivates a marker that defines a fixed reference point
for relative marker analysis.
If necessary, the command activates a marker and positions it on the peak power.
Subsequently, you can change the coordinates of the fixed reference independent of
the marker. The fixed reference is independent of the trace and is applied to all active
delta markers.
Suffix:
<n>
<m>
Parameters:
<State>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
Window
Marker
Numeric value that defines the vertical position of the reference.
The unit and value range is variable.
*RST:
Fixed Reference: OFF
Default unit: DBM
CALC:DELT:FUNC:FIX:RPO:Y -10dBm
Sets the reference point level for delta markers to -10 dBm.
See
"Defining a Fixed Reference"
See
"Defining Reference Point"
Window
Marker
Numeric value
*RST:
0
Default unit: dB
Window
Marker
ON | OFF | 0 | 1
Remote commands
on page 362
on page 379