The error message remains in place when you switch to "Local" mode. To close the
message box, select the
Only the most recent error is displayed in remote mode. However, in local mode, all
errors that occurred during remote operation are listed in a separate tab of the "Net-
work + Remote" dialog box (see
Remote command:
SYSTem:ERRor:DISPlay
SYSTem:ERRor:CLEar:REMote
9.4.3
Compatibility settings
The R&S FPL1000 can emulate the GPIB interface of other signal and spectrum ana-
lyzers, e.g. in order to use existing control applications.
Compatibility with former R&S signal and spectrum analyzers
As a rule, the R&S FPL1000 supports most commands from previous R&S signal and
spectrum analyzers such as the FSQ, FSP, FSU, or FSV. However, the default values,
in particular the number of sweep points or particular bandwidths, may vary. Therefore,
the R&S FPL1000 can emulate these other devices, including their default values, in
order to repeat previous measurements or support existing control applications as in
legacy systems.
The required settings are configured in the "Compatibility" tab of the "Network
+Remote" dialog box.
Manuel d'utilisation 1179.5860.09 - 12
Network operation and remote control
Network and remote control settings
"Close" icon.
Chapter 9.4.5, "Remote
on page 974
on page 977
errors",
on page 580).