R&S
FPL1000
DISP:TRAC:Y:RLEV -10dBm
//Activate occupied bandwidth measurement.
CALC:MARK:FUNC:POW:SEL OBW
//Set the percentage of power to 99%.
POW:BWID 99PCT
//Set the channel bandwidth to 21 kHz.
POW:ACH:BAND 21kHz
//Optimize the instrument settings according to the channel bandwidth.
POW:ACH:PRES OBW
//Determine the ideal reference level for the measurement.
POW:ACH:PRES:RLEV
//Set the trace detector to positive peak.
DET APE
//--------------Performing the Measurement-----
//Select single sweep mode.INIT:CONT OFF
//Initiate a new measurement and waits until the sweep has finished.
INIT;*WAI
//---------------Retrieving Results-------------
//Return the occupied bandwidth.
CALC:MARK:FUNC:POW:RES? OBW
10.6.7
Measuring the spectrum emission mask
All remote control commands specific to spectrum emission mask measurements are
described here.
See also
Remote commands exclusive to spectrum emission mask measurements:
Managing measurement configurations
Controlling the measurement
Configuring a multi-sem measurement
Configuring a sweep list
Configuring the reference range
Configuring the power classes
Configuring the list evaluation
Performing an SEM measurement
Retrieving results
Example: SEM measurement
10.6.7.1
Managing measurement configurations
The following commands control measurement configurations for SEM measurements.
Manuel d'utilisation 1179.5860.09 - 12
Chapter 10.6.2, "Configuring power
................................................................................................ 655
............................................................................................................ 677
Configuring and performing measurements
measurements",
...................................................................... 651
....................................................................................... 653
....................................................................... 654
.................................................................................. 669
..................................................................................... 671
...................................................................................... 676
.............................................................................. 677
....................................................................................... 678
Remote commands
on page 621.
657
657

Publicité

loading