[SENSe:]LIST:RANGe<ri>:BREak <State>
This command controls the sweep for all ranges.
Suffix:
<ri>
Parameters:
<State>
Example:
[SENSe:]LIST:RANGe<ri>:COUNt?
This command queries the number of ranges in the sweep list.
Suffix:
<ri>
Return values:
<Ranges>
Example:
Usage:
[SENSe:]LIST:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
you cannot delete the reference range
a minimum of three ranges is mandatory.
Suffix:
<ri>
Example:
[SENSe:]LIST:RANGe<ri>:DETector <Detector>
This command selects the detector for a spurious emission measurement range.
Suffix:
<ri>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
irrelevant
ON | 1
The R&S FPL1000 stops after measuring one range, and the
status bit number 10 in the STAT:OPER register is set.
To continue with the next range, use INITiate<n>:CONMeas.
OFF | 0
The R&S FPL1000 sweeps all ranges in one go.
*RST:
0
LIST:RANG2:BRE ON
irrelevant
Number of ranges in the sweep list.
LIST:RANG:COUN?
Query only
1..n
Selects the measurement range.
LIST:RANG2:DEL
1..n
Selects the measurement range.
Remote commands
689
689