Return values:
<SystemMessages> String containing all messages in the "System Messages" table.
<RemoteErrors>
Usage:
SYSTem:ERRor[:NEXT]?
This command queries the most recent error queue entry and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
For details on error queues see
Usage:
SYSTem:FIRMware:UPDate <Directory>
This command starts a firmware update using the *.msi files in the selected directory.
The default path is D:\FW_UPDATE. The path is changed via the
command. To store the update files the
Only user accounts with administrator rights can perform a firmware update.
Setting parameters:
<Directory>
Example:
SYSTem:FORMat:IDENt <IDNFormat>
This command selects the response format to the
Manuel d'utilisation 1179.5860.09 - 12
REMote
Queries the error messages that occurred during remote opera-
tion.
Note: The remote error list is automatically cleared when the
R&S FPL1000 is shut down.
<Error_no> | <Description> | <Command> | <Date> | <Time>
Comma-separated list of errors from the "Remote Errors" table,
where:
<Error_no>: device-specific error code
<Description>: brief description of the error
<Command>: remote command causing the error
<Date>|<Time>: date and time the error occurred
Query only
Chapter 9.2, "Status reporting
Query only
MMEMory:DATA
SYST:FIRM:UPD 'D:\FW_UPDATE'
Starts the firmware update from directory "D:\FW_UPDATE".
Remote commands
Configuring the R&S FPL1000
system",
on page 557.
MMEMory:COMMent
command is used.
*IDN?
query.
985