<GateLength>
[SENSe:]LIST:POWer:STATe <State>
This command turns the List Evaluation off.
Parameters:
<State>
10.6.15.2
Example: performing list evaluation
The following example shows a list evaluation with the following configuration.
No
1
2
3
-----Measurement with synchronization via service request-----
*ESE 1
*SRE 32
// Configures the status reporting system to produce a service request.
LIST:POW:SET ON,ON,OFF,EXT,POS,10us,434us
//Turns on the list evaluation, configures the global list evaluation settings and
//evaluates the peak and RMS power.
LIST:POW
935.2MHZ,0,10,OFF,NORM,1MHZ,3MHZ,440us,0,
935.4MHZ,0,10,10,CFIL,30KHZ,100KHZ,440us,0,
935.6MHZ,0,10,20,CFIL,30KHZ,100KHZ,440us,0;
*OPC
//Defines a list with 3 entries and initiates the measurement with synchronization to the end
//Analyzer produces a service request
//On service request:
SENS:LIST:POW:RES?
//Returns the results of the measurements, two for each frequency (peak and RMS power).
-----Initiliazing the measurement and querying results simultaneously-----
Manuel d'utilisation 1179.5860.09 - 12
Defines the gate length for gated measurements.
Setting 0 seconds turns gated measurements off.
To perform gated measurements, the trigger source must be dif-
ferent from IMMediate.
Range:
*RST:
Default unit: s
OFF | 0
*RST:
Freq
Ref
RF
[MHz]
Level
Attenu-
[dBm]
ation
[dB]
935.2
0
10
935.4
0
10
935.6
0
10
Configuring and performing measurements
31.25 ns to 30 s
0 s
0
El
Filter
RBW
Attenu-
ation
[dB]
---
Normal
1 MHz
10
Channel
30 kHz
20
Channel
30 kHz
Remote commands
VBW
Meas
Trigger
Time
Level
3 MHz
440 µs
0
100 kHz
440 µs
0
100 kHz
440 µs
0
741
741