R&S
FPL1000
Example:
Usage:
[SENSe:]CORRection:TRANsducer:COMMent <Comment>
This command defines the comment for the selected transducer factor.
Before you can use the command, you have to select and turn on a transducer.
Parameters:
<Comment>
Manual operation:
[SENSe:]CORRection:TRANsducer:DATA {<Frequency>, <Level>}...
This command configures transducer factors for specific trace points. A set of trans-
ducer factors defines an interpolated transducer line and can be stored on the instru-
ment.
Parameters:
<Frequency>
<Level>
Example:
Manual operation:
Frequency
0 Hz
2 GHz
4 GHz
[SENSe:]CORRection:TRANsducer:DELete
This command deletes the currently selected transducer factor.
Before you can use the command, you have to select a transducer.
Example:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
SENSE:CORR:TRAN:CAT?
//Result: 2743,2312620544,720,'FactorGSM.TDF',2023,'FactorBTS.TDF'
Query only
*RST:
(empty comment)
See
"Comment"
on page 542
The unit for <Frequency> is Hz, which may or may not be omit-
ted. Frequencies have to be sorted in ascending order.
Default unit: Hz
The unit for <Level> depends
TRANsducer:UNIT
SENSe1:CORRection:TRANsducer:UNIT 'DB'
// Frequency Span 0 Hz to 4 Ghz
SENSe1:CORRection:TRANsducer:DATA 0,8,2GHz,5,4GHz,3
Creates the transducer points:
See
"Data Points"
on page 542
Level
8 dB
5 dB
3 dB
CORR:TRAN:DEL
See
"Delete Line"
on page 541
Remote commands
Configuring the R&S FPL1000
on[SENSe:]CORRection:
.
970
970