R&S
FPL1000
<m>
Parameters:
<State>
Example:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BWIDth <Bandwidth>
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BANDwidth <Bandwidth>
This command defines the bandwidth around the center frequency that is included in
the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Suffix:
<n>
<m>
Parameters:
<Bandwidth>
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold <Level>
This command defines the threshold level for the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Suffix:
<n>
<m>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
irrelevant
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
//Activate signal tracking to keep the center frequency on the signal pea
//After each sweep the maximum on trace 1 is searched within a range of 2
//around the center frequency. It must have a minimum power of -90dBm.
CALC:MARK:FUNC:STR ON
CALC:MARK:FUNC:STR:BAND 20MHz
CALC:MARK:FUNC:STR:THR -90dBm
CALC:MARK:FUNC:STR:TRAC 1
See
"Signal Tracking"
on page 314
irrelevant
irrelevant
Range:
10 Hz to Max span
*RST:
(= span/10 on activating the function)
Default unit: Hz
See
"Signal Tracking"
on page 314
irrelevant
irrelevant
Remote commands
768
768