Restoring the connection to the R&S FPL1000
Follow the instructions above for setting up a connection to the R&S FPL1000. If the
connection is terminated and then restored, the R&S FPL1000 remains in the same
state.
9.5.7.4
How to shut down the R&S FPL1000 via remote operation
1. Select the R&S FPL1000 softfrontpanel and close the application with the key com-
bination [ALT] + [F4].
2. Select the desktop and press the key combination [ALT] + [F4].
A safety query is displayed to warn you that the instrument cannot be reactivated
via remote operation and asks you whether you want to continue the shutdown
process.
3. Respond to the safety query with "Yes".
The connection with the controller is terminated and the R&S FPL1000 is shut
down.
9.5.8
How to start a remote control session from a PC
When you switch on the R&S FPL1000, it is always in manual operation state ("local"
state) and can be operated via the front panel.
To start remote control
1 .
Send an addressed command (GTR - Go to Remote) from a controller to the instru-
ment.
The instrument is switched to remote control ("remote" state). Operation via the
front panel is disabled. Only the "Local" softkey is displayed to return to manual
operation. The instrument remains in the remote state until it is reset to the manual
state via the instrument or via remote control interfaces. Switching from manual
operation to remote control and vice versa does not affect the other instrument set-
tings.
During program execution, send the SYSTem:DISPlay:UPDate ON command to
2 .
activate the display of results (see
The changes in the device settings and the recorded measurement values are dis-
played on the instrument screen.
3 .
To obtain optimum performance during remote control, send the
SYSTem:DISPlay:UPDate OFF command to hide the display of results and dia-
grams again (default setting in remote control).
4 .
To prevent unintentional return to manual operation, disable the keys of the instru-
ment using the universal command LLO.
Manuel d'utilisation 1179.5860.09 - 12
Network operation and remote control
How to set up a network and remote control
SYSTem:DISPlay:UPDate
on page 974).
607
607