Les langues disponibles

®
R&S
FPL1000
For details see
Parameters:
<State>
Example:
Manual operation:
FETCh:CMEasurement:P1DB[:RESult]?
FETCh:CMEasurement:P3DB[:RESult]?
FETCh:CMEasurement:PNDB[:RESult]?
Queries the results of the n dB compression point measurement, see
CMEasurement:RESult:PNDB[:STAT]
To query the compression value, use
ment:RESult:CVALue?.
Example:
Usage:
Manual operation:
10.6.3.2
Programming example: measuring compression points
This example demonstrates how to determine compression points for an amplifier in a
remote environment.
//-----------Configuring the measurement ------------
//Reset the instrument
*RST
//Configure internal generator for power sweep from -30 dBm to -20 dBm
//with level offset of 2 dB
OUTP ON
SOUR:POW:MODE SWE
SOUR:POW:STAR -30
SOUR:POW:STOP -20
SOUR:POW:OFFS -2
//Activate compression point measurement for 3 dB and 5 dB compression.
CONF:CME ON
CONF:CME:RES:P3DB ON
CONF:CME:RES:CVAL 5
CONF:CME:RES:PNDB ON
Manuel d'utilisation 1179.5860.09 - 12
Chapter 6.2.3, "Power sweep
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
CONF:CME ON
See
"Compression Measurement State"
CONFigure:CMEasure
FETC:CME:PDNB?
//Result: -3 dBm, 3 dBm
Query only
See
"Compression Points"
Remote commands
Configuring and performing measurements