Manual Vent Valve
The Manual Vent Valve is used to vent the high-pressure gas from the Visual-ICE Cryoablation System if the automatic
venting feature is not used�
Needle Connection Panel
Figure 4. Visual-ICE Cryoablation System Needle Connection Panel
Figure 5. Visual-ICE Cryoablation System Needle Channel
1 MTS Connection Port
2 Software Reset
MTS Connection Ports
Four MTS Connection Ports are located above the Needle Connection Channels� The MTS Connection Ports support
connection of the Boston Scientific Multi-Point 1�5 Thermal Sensor Devices that are used to monitor temperature
within and near the cryoablation target site�
Software Reset
The Software Reset button is used to boot the Visual-ICE Cryoablation System into a recovery mode if the software
becomes corrupted (reference the Software Recovery Section)�
3 Power Control Knob
5 Locking Bar
4 Needle Channel
6 Needle Port with Electrical Connection
7 Channel Number
9