Boston Scientific Visual-ICE Manuel D'utilisation page 80

Systeme de cryoablation

Publicité

Message
20-07
Stop All Channels
• You selected to control all needles simultaneously
• Do you want to stop activity in all needles now?
YES
20-10
Cautery
• CAUTION: Activation of thermal cautery immediately
ceases all other cryoablation operations for the duration of
cautery activity�
• Do you want to proceed with cautery?
YES
20-11
End Procedure
• Are you sure you want to end the procedure?
YES
20-12
Automatic Vent
• Do you want to automatically vent high pressure gas from
the system?
YES
20-13
Automatic Vent
• Is the gas supply closed?
YES
20-14
Automatic Vent
• The gas pressure is not dropping
• Check that the gas cylinder shut off valve is closed
20-15
Automatic Vent
• Venting is in progress
• If needles are still connected, do not unlock channels or
disconnect needles until venting is complete
20-16
Automatic Vent
• Automatic venting successfully completed
20-17
Gas Vent
• Before disconnecting the gas hose, manually vent the
Visual-ICE Cryoablation System using the Manual Vent Valve
on the rear of the machine
20-18
System Shutdown
• Do you want to shut down the system?
YES
Black (K) ∆E ≤5.0
NO
NO
NO
NO
CANCEL
NO
Reason for Appearance / Solutions
User selected Stop button in ALL channel to control
all active channels� To proceed, the user must
confirm stop for all active needles�
User pressed Thaw button to open the Advanced
Thaw Controls and the thermal cautery option�
The CAUTION is displayed upon user selecting
advanced option for thermal cautery�
User selected End Procedure and must confirm
desire to end the procedure�
User was provided an option for automatically
venting the high pressure gas in the system�
If user selected the option for automatic gas
venting, the user will need to confirm the gas supply
is closed before automatic venting is activated�
User selected to automatically vent high pressure
gas at the end of a procedure, but the pressure was
not dropping� User must ensure the shut off valve
is closed�
User selected to automatically vent the high
pressure gas at the end of the procedure�
User chose to automatically vent the high pressure
gas at the end of the procedure�
User selected to not use the feature to
automatically vent high pressure gas from the
System�
User selected Shutdown in the login screen to
power down the system�
80