Message
20-19
Procedure Timeout
• The procedure has exceeded the allowable time
• Procedure will be terminated
20-20
Maximum Activity for FastThaw Function
• Maximum limit for FastThaw Function reached
• Wait until thawing is complete before activating additional
FastThaw Function capable needles
• This channel will use passive thaw
20-21
Maximum Activity for i-Thaw Function
• Maximum limit for i-Thaw Function reached
• Wait until thawing is complete before activating additional
i-Thaw Function capable needles
• This channel will use passive thaw
20-22
Maximum Limit for Active Thaw
• Maximum limit for i-Thaw Function and FastThaw Function
reached
• Wait until thawing is complete before activating thaw on
additional needles
20-23
Maximum Activity for Test
• Maximum limit for test reached
• Wait until testing is complete before activating additional
needles for testing
20-24
Advanced Thaw Unavailable
• Needle(s) in channel X are not capable of advanced thaw
• Advanced thaw activity is unavailable
20-25
Advanced Thaw Unavailable
• Advanced thaw activity is unavailable on channel X while
helium is connected
• Use helium thaw or disconnect helium
Reason for Appearance / Solutions
The procedure exceeded the allowable 8 hour
duration�
User attempted to activate more than four
FastThaw Function capable needles� The Visual-
ICE Cryoablation System supports simultaneous
operation of no more than four FastThaw Function
capable needles� After completing the thaw phase
with four needles, additional needles may be
activated for FastThaw Function�
Thawing with i-Thaw Function is limited to a
maximum of 9 needles simultaneously activated in
thaw mode�
When thawing with CX-type needles, thawing is
limited to a maximum of 7 needles simultaneously
activated�
After completing the thaw phase, additional
needles may be activated�
User attempted to activate more needles than are
supported for active thaw� After completing the
thaw phase, additional needles may be activated�
User attempted to initiate testing on more needles
than are supported for simultaneous testing�
User attempted to initiate Advanced Thaw in
a channel� The channel did not contain i-Thaw
Function or FastThaw Function capable needles�
User attempted to initiate Advanced Thaw while
helium was connected� If helium is connected,
the system defaults to helium thaw mode� If
mixed needle types are connected (some helium
thaw and some i-Thaw Function), the system also
defaults to helium thaw mode�