Pre-Procedure Testing
WARNING: Prior to starting a cryoablation procedure, set up the Visual-ICE Cryoablation System and perform
Needle Integrity and Functionality Tests on each cryoablation needle and thermal sensor.
1� On the touch screen monitor, press Start Procedure� The Procedure Screen appears (Screen 11)�
Screen 11. Procedure Screen
2� Using aseptic technique, carefully remove the cryoablation needle from the package and place in a sterile
work area�
3� Remove the connector cap, and then connect the needle to the Visual-ICE Cryoablation System needle
connection panel (Figure 4)�
WARNING: Do not kink, pinch, cut or pull excessively on the needle tubing� Damage to needle handle or tubing may
cause the needle to become unusable�
4� After a needle(s) is inserted into the desired channel, lock the channel by sliding the locking bar away from the
center of the system (Figure 9)�