Systeme de cryoablation

Publicité

SYSTEM
Message
80-01
Communication Failure
• Internal communication failed
• Reconnection attempt failed
• Restarting Visual-ICE Cryoablation System
• If problem persists, contact Boston Scientific
Technical Assistance Center
80-02
Startup Failure
• System self-checks failed
• Restart the system
• If problem persists, contact Boston Scientific
Technical Assistance Center
80-03
Pressure Alert
• Pressure exceeds safe operating limits
• Close the gas cylinders
• Procedure will be terminated and gas vented
from the system
80-04
Temperature Warning
• The internal temperature of the Visual-ICE
Cryoablation System exceeds appropriate
operating limits
• Discontinue the cryoablation procedure as
soon as safe to do so
• Contact Boston Scientific Technical Assistance
Center
80-05
Service Due
• Low battery detected
• Contact Boston Scientific Technical Assistance
Center to schedule service
80-30
System Error*
• Gas pressure display may be inaccurate�
Monitor procedure carefully using image
guidance� Use passive thaw� At the conclusion
of the procedure, contact Boston Scientific
Technical Assistance Center�
80-31
System Error*
• Gas cylinder shut off valve is not open
enough to provide adequate flow� If necessary,
open the valve approximately another half
turn�
80-32
System Error*
• Channel X is defective� Choose another
channel�
At the conclusion of the procedure, contact
Boston Scientific Technical Assistance Center�
Reason for Appearance / Solutions
Software was unable to connect with hardware after attempt to
reinitialize communication� If the restart fails, the system is not
usable�
Software self-checks located a failure that required the system to
be restarted�
System detected the internal pressure was above safe limits�
The system will terminate the procedure and vent gas from the
system�
The internal temperature of the system exceeded appropriate
operating limits�
System detected a low battery� System operation may be
impacted if operated with a low battery�
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
Internal pressure checks were inconsistent and may produce an
inaccurate display on the Pressure Gauge�
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
Gas flow from gas cylinder was insufficient� To improve gas flow,
the cylinder valve should be opened further�
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
A defective solenoid on Channel X was detected; another
channel should be selected�