Message
80-33
System Error*
• Defective MTS in channel [X]� Replace with a
new MTS�
80-34
System Error*
• Fan X failure� At the conclusion of the
procedure, contact Boston Scientific Technical
Assistance Center�
80-35
System Error*
• The gas flow rate in channel XX exceeds
the system operation requirements and may
impact performance� Reduce the number of
active needles�
80-36
System Error*
• The gas flow rate exceeds the system
operation requirements and may impact
performance� Reduce the number of active
needles�
80-37
System Error*
• Defective MTS in channel 1, 2, 3, 4
80-38
System Error*
• Passive thaw required
SERVICE
Message
90-01
Service Due
• Service the Visual-ICE Cryoablation System soon
• Service must be completed by [DATE]�
• Contact Boston Scientific Technical Assistance Center to schedule
service�
90-02
Service Due
• Service for the Visual-ICE Cryoablation System is past due
• Contact Boston Scientific Technical Assistance Center to schedule
service�
90-03
System End of Life
• The Visual-ICE Cryoablation System is at the end of operational life
• Contact Boston Scientific Technical Assistance Center to arrange
return of this system to be refurbished, replaced or disposed�
Black (K) ∆E ≤5.0
Reason for Appearance / Solutions
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
The MTS temperature display was out of the expected range when
initially connected to the needle connection panel�
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
Fan X was identified as not operating�
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
The calculated flow rate for a given channel exceeded the
optimum system requirement� The number of active needles
should be reduced�
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
The cumulative flow rate for all channels exceeded the optimum
system requirement� The number of active needles should be
reduced�
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
Contact Boston Scientific Technical Assistance Center�
* System Error messages are displayed in the right corner of the
navigation tool bar� The User is offered details of the error by
"Press here" instruction�
Contact Boston Scientific Technical Assistance Center�
Reason for Appearance / Solutions
User was reminded of the deadline to
complete scheduled system service�
Reminder begins four weeks prior to
service due date�
The Visual-ICE Cryoablation System
was not serviced on schedule� During
subsequent startups, the User is
reminded the service is past due�
The Visual-ICE Cryoablation System
reached the end of its operating life�
Contact Boston Scientific Technical
Assistance Center to schedule service�