TEMPERATURE SENSOR / ORGAN MAP
Message
50-01
Alarm
• Sensor point (5, 15, 25 or 35) on MTS (#X) met selected alarm limit
50-02
Alarm
• Sensor point (5, 15, 25 or 35) on MTS (#X) met selected alarm limit
50-03
MTS Disconnected
• An MTS is disconnected
• Reconnect the MTS to continue to monitor temperature in that
location
50-05
Clear Drawing
• Are you sure you want to erase the entire drawing?
YES
NO
50-06
Organ Type
• All annotation and needles will be removed from the organ
before proceeding
• Would you like to proceed with the organ type change?
YES
NO
SOFTWARE
Message
60-04
Update Failed
• An error occurred during the update process
• Retry update
60-04
Update Failed
• An error occurred during the upload process
• Retry upload
60-05
Incompatible Hardware
• The hardware is not compatible with the current
software
• Contact Boston Scientific Technical Assistance Center to
schedule service
60-06
Incompatible Software
• The Visual-ICE Cryoablation System software is not
compatible with regulatory approvals
• Contact Boston Scientific Technical Assistance Center
to schedule service
60-07
Invalid Configuration
• A software file is nonoperational�
• If problem persists, contact Boston Scientific
Technical Assistance Center
60-09
Software Recovery
• This will rollback the software on this system
• Are you sure you want to do this?
YES
NO
Reason for Appearance / Solutions
The MTS detected the selected alarm
temperature�
The MTS detected the selected alarm rate of
temperature change within 30 seconds�
An MTS became disconnected during a
procedure�
User selected the Clear Drawing button�
When selected, all information placed on
the organ map, except needle placement,
will be erased�
User changed the organ type�
Reason for Appearance / Solutions
During User update of the system, an error occurred that
prevented completion of the update� Another update
should be attempted�
During User update of the system, an error occurred that
prevented completion of the upload� Another upload
should be attempted�
When the system performed self-checks at start up, an
incompatibility between the hardware and software was
detected� Contact Boston Scientific Technical Assistance
Center�
The software was checked against approved software
versions in the regulatory files for each respective market�
Incompatibility with regulatory approval was identified� The
system should be updated with the appropriate software�
Contact Boston Scientific Technical Assistance Center�
There was a problem with the software's configuration
files� Contact Boston Scientific Technical Assistance
Center�
User pressed the Software Recovery button then
selected Rollback� Activation will rollback the software to
the previous software version�
83