• Similarly, when FastThaw Function is selected, pressing the Change to i-Thaw Function button will change the
thaw from FastThaw Function to i-Thaw Function (Screen 34)�
Screen 34. Advanced Thaw Controls i-Thaw Function
2� After selecting the type of thaw, press the Thaw or FastThaw Function button in the Channel Control to initiate
thawing�
NOTE: The FastThaw Function generates a temperature that is higher than that generated for i-Thaw Function,
resulting in a faster thaw time�
NOTE: A maximum of 7 needles can thaw simultaneously using the i-Thaw Function; a maximum of 4 needles can
thaw simultaneously using the FastThaw Function�
3� During the needle warming phase for CX-type needles, the channel status will display a rotating warming
indicator (Screen 35)�
Screen 35. Advanced Thaw - Needle Warming
4� When the threshold temperature for i-Thaw Function or FastThaw Function is reached, the channel status will
display the estimated temperature range for the needle shaft (Screen 36)�
NOTE: The shaft temperature is displayed as a temperature range because tissue and procedural variables will affect
the temperature�