Boston Scientific Visual-ICE Manuel D'utilisation page 8

Systeme de cryoablation

Publicité

Touch Screen Monitor
The cryoablation procedure is controlled through the use of the touch screen monitor� The monitor can be tilted and
rotated to provide the optimal viewing and operating angle for the user� The touch screen monitor includes a virtual
on-screen English QWERTY keyboard for entering procedure related data and may be operated by finger touch� The
monitor folds flat into the monitor storage basin on the top of the unit for system storage�
Communication Ports
2 communication ports are located on the back panel of the monitor storage basin (Figure 3)�
• The Ethernet port is inactive
• The USB 2�0 port allows you to save reports to a USB flash drive for download to another computer or for printing�
Figure 3. Monitor Storage Basin
1
Ethernet Port (Inactive)
2
USB 2�0 Port
Mouse Track Pad
The mouse track pad is located within the recessed monitor storage basin� The mouse track pad provides an alternative
to the touch screen as a means of interacting with the system� Use the track pad to move and position the cursor on the
monitor� To press a button on the screen, position the cursor over the button and press the left button on the track pad�
Storage Compartment
The storage compartment may be used to store Visual-ICE Cryoablation System accessories, such as the gas supply
lines and tools� Do not place very heavy objects in the storage compartment� The weight limit is 23 kg (50 lb)� Do
not store liquids in the compartment� Liquids spilled in the storage compartment may drip into the system; the
compartment is not watertight�
Brake Pedal
The brake pedal operates on the front two wheels of the Visual-ICE Cryoablation System� Position the brake in the
UP position to prevent the front two wheels from swiveling during transportation� Position the brake in the DOWN
position to lock the front two wheels into place� When the brake pedal is in the middle position, the front two wheels
are free to roll and swivel� If the flooring is uneven, it may be necessary to lock the two rear wheels in addition to
locking the front wheels� Lock each rear wheel by using the individual locking pedal on each wheel�
Argon Shutoff Valve
The Argon Shutoff Valve is used to turn the gas supply to the Visual-ICE Cryoablation System ON or OFF� It should be
left in the Argon ON position and used to turn OFF argon gas only in an emergency�
Gas Inlets
The gas supply lines connect the argon and helium gas supply from the respective gas cylinders to the argon and
helium gas inlets� The argon inlet is a plug connector; the helium inlet is a socket connector�
Black (K) ∆E ≤5.0
8