Manual Software Update
Administrative and service users can manually update the Visual-ICE Cryoablation System software via a USB flash drive�
1� Press the Manual Software Update button on the Configure Settings Screen (Screen 46)�
2� Press the Browse button to select the update file and press Update� A confirmation message appears when the
software update is complete (Screen 47)�
Screen 47. Software Update Confirmation
POST-PROCEDURE
Any serious incident that occurs in relation to this device should be reported to the manufacturer and to the relevant
local regulatory authority�
For customers in Australia, report any serious incident that occurs in relation to this device to Boston Scientific and to
the Therapeutic Goods Administration (https://www�tga�gov�au)�
Cleaning the Visual-ICE Cryoablation System
Clean the Visual-ICE Cryoablation System after each use by following the steps below�
1� Clean the Touch Screen Monitor when the Visual-ICE Cryoablation System is powered OFF�
• Wipe the screen gently with a damp gauze wipe�
• Use water or isopropyl alcohol cleaning solutions�
• Do not use cleaning agents such as Betadine Antiseptic Solution or bleach solution�
2� Clean the System by wiping with a damp gauze wipe�
• Use soap and water or isopropyl alcohol cleaning solutions�
• Do not use cleaning agents such as Betadine Antiseptic Solution or bleach solution�
• Do not allow water or other liquid to drip or seep inside the needle connection ports� The needle connection
ports must remain completely dry at all times�
3� Ensure that cleaned surfaces are dry prior to closing or turning on the system�
Disposal
All external and accessible surfaces of this device should be cleaned per the Cleaning the Visual-ICE Cryoablation
System instructions included in the user manual� Include any common detachable cables (power cord, video cables,
patch cords, etc�)� Review the user manual to identify if any hazardous materials are present�
If placing the unit into an electronics recycling stream, notify the receiver of the presence of such materials� Use
of recycling service suppliers familiar with medical electrical equipment is recommended, but not required� Do not
dispose of by incineration, burial or placement into municipal waste stream�
Device should be safely disposed of in accordance with hospital, administrative, and/or local government policy or
returned to Boston Scientific� For a Returned Product Kit, contact Boston Scientific Technical Assistance Center�
Dispose of all sharps directly into a sharps disposal container labeled with a biological hazard symbol� Sharps waste
should be safely disposed of using available sharps waste channels in accordance with hospital, administrative, and/
or local government policy�