SYSTEM SPECIFICATIONS
Mechanical Specifications
Visual-ICE Cryoablation System
• Weight: 100 kg (220 lb)
• Height: 107 cm (42 in), monitor down
157 cm (62 in), monitor up
• Foot Print: 56 cm x 66 cm (22 in x 26 in)
• Storage Compartment Weight Capacity: 22 kg (50 lb)
• Monitor Storage Basin Weight Capacity: 9 kg (20 lb)
• Closed Monitor Weight Capacity: 9 kg (20 lb)
External Gas Supply
• Argon Cylinder:
o Purity Level: 99�998 % or higher
o Solid particle size: < 5 μm
• Helium Cylinder:
o Purity Level: 99�995 % or higher
o Solid particle size: < 5 μm
Gas Cylinder Specifications
• Maximum Pressure: 6000 psi (414 bar, 41�4 MPa)
• Recommended Volume of Gas Cylinder: 42 L – 50 L
Accuracy of Displayed Values
• Temperature Accuracy:
o ± 3 °C over range of -60 °C to 40 °C
• Supplied Gas Pressure Accuracy:
o ± 50 psi, over range of 1000 psi to 6000 psi
o ± 3�4 bar over range of 69 bar to 414 bar
o ± 0� 344 MPa over range of 6�9 MPa to 41�4 MPa
• Built-in Regulator Gas Pressure:
o ± 50 psi, over range of 1000 psi to 4000 psi
o ± 3�4 bar over range of 69 bar to 276 bar
o ± 0�344 MPa over range of 6�9 MPa to 27�6 MPa
• Time Intervals:
o ± 5 seconds over any 10 minute interval
Essential Performance
The Visual-ICE Cryoablation System essential performance is defined as:
• The ability to turn on and turn off the flow of argon or helium to the connected cryoablation needles via
command inputs from the user interface
• The ability to maintain uninterrupted gas flow to user-selected needles when gas flow is turned on by the user
• The ability to uninterruptedly prevent gas flow to user-selected needles when gas flow is turned off by the user
• The ability to energize the electrical resistance heater on user-selected i-Thaw Function/cautery function
capable needles
• The ability to continuously monitor and display temperatures measured by user-selected thermal sensors
87