PROCEDURE
Message
20-01
Cannot Start Test
• The gas pressure is too low/high to begin a procedure
• Check that the gas cylinders have enough pressure to start
the procedure
20-02
Argon Shut Off Valve
• The argon shut off valve may be closed
• Check and open if necessary
20-03
Test All
• You selected to control all needles simultaneously
• Place the needles, individually or in groups, in the basin
such that the full length of the needle shaft is submerged in
the water or saline� As the test cycles run, look at the needles
to ensure that there are no air bubbles, and that a small ice
ball forms during the freeze portion of the sequence�
• Do you want to initiate test in all needles now?
YES
NO
20-03
Test All
• You selected to control all needles simultaneously
• Place the needles, individually or in groups, in the basin
such that the full length of the needle shaft is submerged in
the water or saline� As the test cycles run, look at the needles
to ensure that there are no air bubbles, and that a small ice
ball forms during the freeze portion of the sequence�
• Do you want to initiate test on all channels or just the
untested channels?
YES
NO
20-04
Freeze All Channels
• You selected to control all needles simultaneously
• Do you want to initiate freeze for all active needles now?
YES
NO
20-05
Freeze All Intensity
• You selected to control all needles simultaneously�
• Do you want [x%] freeze intensity applied to all needles
now?
YES
NO
20-06
Thaw All Channels
• You selected to control all needles simultaneously
• Do you want to initiate thaw for all active needles now?
YES
NO
Reason for Appearance / Solutions
User pressed the Test button when either gas
cylinders were not yet connected or the gas
pressure is below operating pressure (Reference
Table 7)� To proceed, gas cylinders with sufficient
pressure must be connected�
At system start-up, the system detected gas
connected, but no gas was entering the system�
The argon shut off valve may be closed� To
proceed, the argon shut off valve must be open�
User selected Test button in ALL channel to control
all active channels� To proceed, the user must
confirm test for all connected needles�
User selected Test button in ALL channel to control
all active channels� To proceed, the user must
confirm test for all connected needles or just the
untested channels�
User selected Freeze button in ALL channel to
control all active channels� To proceed, the user
must confirm freezing of all active needles�
User selected a freeze intensity in ALL channel to
control all active channels� To proceed, the user
must confirm the selected intensity is to be applied
to all active needles�
User selected Thaw button in ALL channel to
control all active channels� To proceed, the user
must confirm thawing of all active needles�
79
Black (K) ∆E ≤5.0