CONFORMANCE TO STANDARDS
Electrical Specifications:
• Input Voltage: 100 VAC to 240 VAC, single phase
• Input Frequency: 50 Hz – 60 Hz
• VA Rating: 250 VA
• IP Rating: IP10
• Fuse Rating: T 3�15AL
• Electrical Protection: Class I, Type BF protection against shock
• Signal Input/Output Ports: one (1) Ethernet port (inactive), one (1) USB 2�0 port
Electromagnetic Compatibility and Immunity (EMC and EMI)
The Visual-ICE Cryoablation System needs special precautions regarding electromagnetic compatibility (EMC) and
needs to be installed and placed into service according to the EMC information provided below�
The Visual-ICE Cryoablation System has been tested in the operating room environment for electromagnetic
compatibility (EMC) and electromagnetic interference (EMI) compliance� The Visual-ICE Cryoablation System has been
tested to be in compliance with IEC 60601-1-2 and EN 55011�
Portable and mobile radiofrequency (RF) communications equipment can affect the Visual-ICE Cryoablation System,
causing it to operate improperly�
Table 1. Cable Lengths
Cable
Power Cable
Gas Tubing (connected to needles)
Gas Supply Line (connected to argon cylinder)
NOTE: The Gas Supply Line is available in more than one length to accommodate procedure room variations�
WARNING: Use of cables other than those specified, with the exception of those sold by Boston Scientific for use
as replacement parts for internal components, may result in increased emissions or decreased immunity of the
Visual-ICE Cryoablation System�
WARNING: The Visual-ICE Cryoablation System should not be used adjacent to or stacked with other equipment�
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 in) to any part of the Visual-ICE Cryoablation System, including
cables specified for use with the system� Otherwise, degradation of the performance of this equipment could result�
Black (K) ∆E ≤5.0
Length
4�6 m (15 ft)
2�5 m (8 ft)
Available lengths: 5 m (16 ft), 8 m (26 ft)