ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems

Impedances Out Of Range Screen

The impedances must be within the acceptable range before MRI Mode is enabled. If the impedances are not within the
acceptable range, the Remote Control will display an error message.
1. Press
to continue.
2. The Remote Control displays a new message instructing the user to review the MRI scan risks related to abnormal
impedances. Review the "Impedances Out of Range" section under the "Safety Information" section of this manual
before proceeding. Press
3. Select
Yes to proceed with enabling MRI Mode or
Warning: An MRI scan is not recommended when the impedances are not within the acceptable range. Higher or
lower than normal impedances could indicate compromised Stimulator‑Lead integrity. Scanning under
these conditions may increase the risk of potential adverse effects that are listed in the "Safety Information"
section of this manual under "Potential Interactions with MRI Environment."
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
92328636-03
30 of 957
to continue.
No to cancel the action.