Bionic Navigator™ 3D 4.0 Programming Manual

Features of the Options Screen

Depending on the Stimulator type that the CP is connected to, the Options screen displayed in this
manual may look different and some options will not be available.
MicroBurst*
A burst waveform with increased resolution in the 0 to 1 second range, within and
between packets of stimulation.
Continuous
Constant stimulation.
This is the default stimulation state for all Programs and Areas.
Cycle
Specifies the on and off duration of stimulation delivery for an entire Program or for
individual Areas within a Program. Cycling is often used to conserve energy.
*Microburst Settings result in the creation of bursts from 2 to over 1,000 pulses.
The Maximum and Minimum Amplitude options, or Max and Min mA, limits the stimulation amplitude.
Cycling and Max mA can be specified for an entire Program, or for individual Areas within the Program.
Ramp Up is the amount of time to gradually increase stimulation from zero to the programmed
amplitude. Ramp Up time can be specified for Areas or Programs.
Figure 24: Program Options
Bionic Navigator™ 3D 4.0 Programming Manual
92395591-02 42 of 1595