MRI Mode for Full Systems
Charge Stimulator Now or Disable MRI Mode Screen (Rechargeable Stimulators Only)
If MRI Mode has already been enabled and the Stimulator battery power falls below the recommended value, the
Remote Control will display a message instructing the patient to charge the Stimulator.
To charge the Stimulator without disabling MRI Mode:
1. Do not press
.
2. Instruct the patient to charge the Stimulator.
3. Check the Remote Control to confirm that the error message has cleared.
4. Navigate to the Home screen on the Remote Control by pressing
on the side panel of the Remote Control and
confirm that the MR Conditional Symbol
is displayed on the Home screen.
The patient can also disable the MRI Mode before charging the Stimulator:
1. Press
to disable the MRI Mode.
2. Instruct the patient to fully charge the Stimulator.
3. Check the Remote Control to confirm that the error message has cleared.
4. Enable MRI Mode by following instructions in the "Enabling MRI Mode" section of this manual.
Caution:
The Charger and Remote Control are MR Unsafe and must not be brought into the MRI scanner room.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
92328636-03
27 of 957