ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems

Vercise Genus Full System

MRI Eligible
The Vercise Genus Stimulator must be implanted under
the skin in a location near the clavicle (pectoral region) or
in the abdomen.
The Vercise Genus Stimulator must be implanted on the
same side of the body as the implanted Lead(s) and Lead
Extension(s).
Patient has up to two Leads implanted.
For a bilateral implant where two Leads and Lead
Extensions are connected to a single Stimulator, both Lead
Extensions are routed on the same side of the body as the
Stimulator.
Extensions directly connected to Stimulator. No adapters
present.
Unused Stimulator Ports have a Port Plug inserted.
Unused Lead Extension Ports when using the 2x8 Contact
Lead Extensions, DB‑3128, have a Port Plug inserted.
The 8 Contact Lead Extension, NM‑3138‑55, has a Lead
inserted (does NOT have a Port Plug inserted).
No Evidence of fractured Leads or compromised
Stimulator‑Lead system integrity.
Single Stimulator is implanted.

Vercise Gevia Full System

MRI Eligible
The Vercise Gevia Stimulator must be implanted under the
skin in a location near the clavicle (pectoral region).
The Vercise Gevia Stimulator must be implanted on the
same side of the body as the implanted Lead(s) and Lead
Extension(s).
Patient has up to two Leads implanted.
For a bilateral implant where two Leads and Lead
Extensions are connected to a single Stimulator, both Lead
Extensions are routed on the same side of the body as the
Stimulator.
Extensions directly connected to Stimulator. No adapters
present.
Unused Stimulator Ports have a Port Plug inserted.
The 8 Contact Lead Extension, NM‑3138‑55, has a Lead
inserted (does NOT have a Port Plug inserted).
No Evidence of fractured Leads or compromised
Stimulator‑Lead system integrity.
Single Stimulator is implanted.
When Vercise Gevia is implanted, DB‑2201 or DB‑2202
Leads with NM‑3138‑55 Lead Extensions are implanted.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
92328636-03
38 of 957
Not MRI Eligible
The Vercise Genus Stimulator is NOT implanted near the
clavicle (pectoral region) NOR implanted in the abdomen.
The Vercise Genus Stimulator is NOT implanted on the
same side of the body as the implanted Lead(s) and Lead
Extension(s).
Patient has more than two Leads implanted.
For a bilateral implant where two Leads and Lead
Extensions are connected to a single Stimulator, both Lead
Extensions are NOT routed on the same side of the body
as the Stimulator.
Extension NOT directly connected to Stimulator. Adapter
is present.
Unused Stimulator Ports do NOT have a Port Plug
inserted.
Unused 2x8 Contact Lead Extension (DB‑3128) Lead
Extension Ports do NOT have a Port Plug inserted.
The 8 Contact Lead Extension, NM‑3138‑55, does NOT
have a Lead inserted (HAS a Port Plug inserted).
Evidence of fractured Leads or compromised
Stimulator‑Lead system integrity.
More than one Stimulator is implanted.
Not MRI Eligible
The Vercise Gevia Stimulator is NOT implanted near the
clavicle (pectoral region).
The Vercise Gevia Stimulator is NOT implanted on the
same side of the body as the implanted Lead(s) and Lead
Extension(s).
Patient has more than two Leads implanted.
For a bilateral implant where two Leads and Lead
Extensions are connected to a single Stimulator, both Lead
Extensions are NOT routed on the same side of the body
as the Stimulator.
Extension NOT directly connected to Stimulator. Adapter
is present.
Unused Stimulator Ports do NOT have a Port Plug
inserted.
The 8 Contact Lead Extension, NM‑3138‑55, does NOT
have a Lead inserted (HAS a Port Plug inserted).
Evidence of fractured Leads or compromised
Stimulator‑Lead system integrity.
More than one Stimulator is implanted.
When Vercise Gevia is implanted, DB‑3128 Lead
Extensions are implanted.