System Configuration

Table 3. Lead‑Only System Configuration
Lead(s)-Only,
Fully Implanted Configuration
Lead(s)-Only,
Externalized Configuration
2
The system has only been evaluated with a Lead Boot. Failure to use a Lead Boot could increase the chance of the risks described in the "Safety
Information" section of this manual under "Potential Interactions with MRI Environment."
• A fully implanted Lead‑Only DBS System is a booted Lead System comprised of Lead(s), Lead
Boot(s), and a Burr Hole Cover or alternative method of securing Leads (see Table 2 for scan
eligible components).
• Fully implanted Lead(s) (not exposed).
• Leads are capped with Lead Boots on proximal ends and excess Lead is coiled and implanted
under the scalp on the skull.
2
• Patients with up to two Leads implanted are scan eligible.
• No evidence can be found of fractured Leads.
• No Extensions or Stimulator present.
• Partially implanted Lead(s) extending out of the patient must be straight with no loops.
• The external portion of the partially implanted Lead cannot be in contact with either the patient or
any part of the scanner (see Table 2 for scan eligible components).
• Patients with up to two Leads implanted are scan eligible.
• No evidence can be found of fractured Leads.
• No Extensions or Stimulator present.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
MR Conditional System Description
92328636-03
5 of 957