ERI or EOS Screens During MRI Mode (Non-Rechargeable Stimulators Only)

If MRI Mode has already been enabled and the Stimulator battery power falls below the threshold, the Remote Control
will display a message informing the patient that the Stimulator has entered the Elective Replacement Indicator (ERI)
period or has reached End of Service (EOS) of the device.
The patient can disable MRI Mode:
1. Press
to disable the MRI Mode.
2. Check the Remote Control to confirm that the Stimulator battery error message still appears.
Caution:
The Remote Control is MR Unsafe and must not be brought into the MRI scanner room.
Warning: Do not perform an MRI scan if MRI Mode is not enabled. Scanning under different conditions may result in
patient injury or device malfunction.