C. DBS Implant Configuration and System Integrity (Check all that apply for Lead‑Only System or Full System)
Fully Implanted Lead-Only System
MRI Eligible
□
Stimulator NOT implanted.
□
Lead Extensions NOT implanted.
□
Leads capped with Lead Boot.
□
Lead(s) fully implanted under the scalp on the skull.
□
Patient has up to two Leads implanted.
□
NO evidence of fractured Leads.
Externalized Lead-Only System
MRI Eligible
□
Stimulator NOT implanted.
□
Lead Extensions NOT implanted.
Partially implanted Lead(s) extending out of the patient are
□
straight with no loops.
The external portion of the partially implanted Lead(s) is
□
NOT in contact with either the patient or any part of the
scanner.
□
Patient has up to two Leads implanted.
□
NO evidence of fractured Leads.
Not MRI Eligible
□
Stimulator implanted.
□
Lead Extensions implanted.
□
Leads NOT capped with Lead Boot.
□
Lead(s) NOT fully implanted under the scalp on the skull.
□
Patient has more than two Leads implanted.
□
Evidence of fractured Leads.
Not MRI Eligible
□
Stimulator implanted.
□
Lead Extensions implanted.
Partially implanted Lead(s) extending out of the patient are
□
NOT straight or HAVE loops.
The external portion of the partially implanted Lead(s) is in
□
contact with either the patient or any part of the scanner.
□
Patient has more than two Leads implanted.
□
Evidence of fractured Leads.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
Appendix A: MRI Patient Eligibility Form
92328636-03
37 of 957