C. DBS Implant Configuration and System Integrity (Check all that apply for Lead‑Only System or Full System)

Fully Implanted Lead-Only System

MRI Eligible
Stimulator NOT implanted.
Lead Extensions NOT implanted.
Leads capped with Lead Boot.
Lead(s) fully implanted under the scalp on the skull.
Patient has up to two Leads implanted.
NO evidence of fractured Leads.

Externalized Lead-Only System

MRI Eligible
Stimulator NOT implanted.
Lead Extensions NOT implanted.
Partially implanted Lead(s) extending out of the patient are
straight with no loops.
The external portion of the partially implanted Lead(s) is
NOT in contact with either the patient or any part of the
scanner.
Patient has up to two Leads implanted.
NO evidence of fractured Leads.
Not MRI Eligible
Stimulator implanted.
Lead Extensions implanted.
Leads NOT capped with Lead Boot.
Lead(s) NOT fully implanted under the scalp on the skull.
Patient has more than two Leads implanted.
Evidence of fractured Leads.
Not MRI Eligible
Stimulator implanted.
Lead Extensions implanted.
Partially implanted Lead(s) extending out of the patient are
NOT straight or HAVE loops.
The external portion of the partially implanted Lead(s) is in
contact with either the patient or any part of the scanner.
Patient has more than two Leads implanted.
Evidence of fractured Leads.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
Appendix A: MRI Patient Eligibility Form
92328636-03