ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems

Vercise Genus or Vercise Gevia

Table 8. Conditions for a Vercise Genus or Vercise Gevia Full System Head Scan Using a Head Transmit Coil
MR Conditional
Static Magnet Strength
Scanner Type
Operating Mode
Head Coil
Maximum Spatial Field Gradient
Maximum Gradient Slew Rate
MRI Coil Setup
Zone Indicated
Maximum B1+rms
Maximum SAR
(if B1+rms is not available)
System Programming
Scan Eligible System Components
System Configuration
Exposure Time

Additional Information

Note: Green colored zones are indicative of scan eligible zones per the scan conditions of Table 8.
8
Stimulation is automatically turned OFF when MRI Mode is enabled. See the "MRI Mode for Full Systems" section of this manual for more information on
MRI Mode including instructions for enabling MRI Mode.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
92328636-03
12 of 957
Yes
1.5T
Horizontal field, cylindrical closed‑bore 1.5T scanner
Normal
4000 gauss/cm (40 T/m)
200 T/m/s per axis
• Head transmit/receive coil (RF quadrature only)
• Receive‑only coil: Any type
• Hydrogen/proton imaging only
Head
2.0 µT for a Vercise Genus or Vercise Gevia Full System
• 0.2 W/kg for Vercise Genus
• 0.1 W/kg for Vercise Gevia
• MRI Mode must be enabled on the device prior to performing scan.
• Rechargeable Stimulators must be fully charged prior to the scan.
• For Vercise Genus, see Table 4 on page 6.
• For Vercise Gevia, see Table 5 on page 7.
See Table 6 on page 8.
Cumulative active scan time (with RF On) should be limited to 30 minutes or
less per imaging session. If 30 minutes of active scan time is reached, allow
60 minutes of non‑active time before proceeding.
• Patient must be positioned in supine or prone position during the scan.
• Patient should be in a psychological condition and mental state to be able
to provide immediate feedback of any problems during the examination.