Vercise Gevia Scan Eligible Components
Table 5. Vercise Gevia Full System Scan Eligible Components
Lead(s)
Extension(s)
Stimulator
Accessories
4
An MRI can also be performed safely if, instead of the Burr Hole Cover, a mini plate similar to a Stryker 12 mm titanium mini plate with Stryker titanium
screws is used to secure the DBS Leads to the skull. Boston Scientific has not evaluated the effect of other implanted devices in combination with or in
proximity to the Boston Scientific Systems described in this manual. Boston Scientific does not recommend an MRI scan if other implanted devices are
present.
• DB‑2201‑30‑AC, 30 cm 8 Contact Standard Lead
• DB‑2201‑30‑DC, 30 cm 8 Contact Standard Lead
• DB‑2201‑45‑BC, 45 cm 8 Contact Standard Lead
• DB‑2201‑45‑DC, 45 cm 8 Contact Standard Lead
• DB‑2202‑30, Vercise™ Cartesia™ 30 cm 8 Contact DBS Directional Lead
• DB‑2202‑45, Vercise™ Cartesia™ 45 cm 8 Contact DBS Directional Lead
• NM‑3138‑55, 55 cm 8 Contact Lead Extension
• DB‑1200, Vercise Gevia™ 16 Contact Implantable Pulse Generator
• SureTek Burr Hole Cover(s) (optional if using alternative method of securing Leads
○ Provided in Burr Hole Cover, DB‑4600‑C, and Burr Hole Cover Spares Kit, DB‑4605‑C
• Suture Sleeve(s) (optional)
○ Provided in the Vercise™ Physician's Spare Kit, DB‑2500‑C, and with DBS Leads
• Port Plugs
○ Provided in the Port Plug Spares Kit, SC‑4401, and IPG Kit, DB‑1200
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
MR Conditional System Description
)
4
92328636-03
7 of 957