Full System Scan

Testing has demonstrated that the Vercise Genus and Vercise Gevia Systems (including the Leads, Lead Extensions,
and Stimulator; see the "MR Conditional System Description" section of this manual for the definition of a Full System)
are MR Conditional.
Appendix A has an MRI Patient Eligibility form that may be used by the physician to confirm the patient meets the
DBS System Conditions for MRI Scans as described in this manual. A Pre‑MRI Scan Condition Checklist to determine
whether scan conditions have been met can be found in "Appendix B: Pre‑MRI Scan Condition Checklist". All Conditions
of Use must be met for an MRI scan to be performed.
A patient with this system can be safely scanned in an MR system meeting the conditions outlined in Table 8, Table 9,
Table 10, or Table 11.

Full System Head Scan Using a Head Transmit Coil

An MRI may be safely performed on a patient implanted with a Full System that meets the implant and radiology
conditions listed in this section.
Caution:
Read this manual in its entirety before performing a MRI scan on a patient implanted with any component of
the Boston Scientific DBS System.
A patient with this system can be safely scanned in an MR system meeting the conditions outlined in Table 8.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems