ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
Table 7. Conditions for a Fully Implanted or Externalized Lead-Only System Scan
Head Coil
MR Conditional
Static Magnet Strength
Scanner Type
Operating Mode
Maximum Spatial Field Gradient
Maximum Gradient Slew Rate
MRI Coil Setup
Zone Indicated
Maximum B1+rms
Or
Maximum SAR
(if B1+rms is not available)
Body Coil
System Programming
Scan Eligible System Components
System Configuration
Exposure Time
Additional Information
Note: Green colored zones are indicative of scan eligible zones per the scan conditions of Table 7.
Note: The
icon indicates a Stimulator cannot be implanted when using the scan conditions listed in Table 7.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
92328636-03
10 of 957
Yes
1.5T
Horizontal field, cylindrical closed‑bore 1.5T scanner
Normal
4000 gauss/cm (40 T/m)
200 T/m/s per axis
• Head transmit/receive coil (RF quadrature only)
or
• Body transmit/receive coil (RF quadrature only)
• Receive‑only coil: Any type
• Hydrogen/proton imaging only
Any (Full Body)
2.0 µT
0.1 W/kg
N/A
See Table 2 on page 4.
See Table 3 on page 5.
Cumulative active scan time (with RF On) should be limited to 30 minutes or
less per imaging session. If 30 minutes of active scan time is reached, allow
60 minutes of non‑active time before proceeding.
• Patient must be positioned in supine or prone position during the scan.
• If possible, patient should be in a psychological condition and mental state
to be able to provide immediate feedback of any problems during the
examination.