Quality of Wireless Service

Vercise PC and Vercise Gevia DBS Systems
The Vercise PC and Vercise Gevia DBS Systems use a Half‑Duplex, direct point‑to‑point,
primary‑secondary communication system with the characteristics listed in Table 6:
Table 6: Quality of Wireless Service of the Configure Tab
(Vercise PC and Vercise Gevia)
Boston Scientific
Between the Remote Control
Stimulator Type
Non‑Rechargeable
Stimulator
Rechargeable
Stimulator
Vercise Genus DBS System
The Vercise Genus DBS System use a Half‑Duplex, direct point‑to‑point, primary‑secondary
communication system based on Bluetooth Low Energy technology with the typical communication
ranges listed in Table 7:
Table 7: Quality of Wireless Service of the Configure Tab
(Vercise Genus)
Between the
Remote Control and
Implanted Stimulator
(IPG)
3 meters
(9.8 ft)
Data will be resent if not successfully received on supported devices. Sources of in‑band
high interference may result in slow connection, difficulty when pairing devices, or both. If you
experience any of these, you may need to decrease the distance between the communicating
devices. For information on how to improve connection issues, see the "Troubleshooting Wireless
Coexistence Issues" section of this manual.
Timing for Vercise PC and Vercise Gevia DBS Systems
Once a command is initiated by the user, the system will respond in less than 1.5 seconds.
Timing for Vercise Genus DBS System
When a user initiates a communication session, the system will typically respond in 1 to 6 seconds.
The typical data throughput during an active programming session will be more than 10 kbps.
Typical Range
and Stimulator
22 inches (55.8 cm)
36 inches (91.4 cm)
Typical Range
Between the Clinician
Between the Remote
Control and External
Implanted Stimulator
Trial Stimulator (ETS)
6 meters
(19.6 ft)
Vercise™ Deep Brain Stimulation Systems Information for Prescribers
Electromagnetic Compatibility
Between the Wand and
Stimulator
18 inches (45.7 cm)
33 inches (83.8 cm)
Between the Clinician
Programmer and
Programmer and
(IPG)
Stimulator (ETS)
3 meters
(9.8 ft)
92328649-04
External Trial
6 meters
(19.6 ft)
25 of 149