Bionic Navigator™ 3D 4.0 Programming Manual
Changing a Lead's Orientation on the Lead Placement Panel
From the FluoroSync screen:
1. Select a Lead.
2. Select the Retrograde check box to flip the Lead 180 degrees, so that the distal end is
pointing caudally and the proximal tails emerge from the rostral end of the Lead.
3. If desired, uncheck the Retrograde check box to flip the Lead back to its original (default)
position.
Figure 12: Changing Lead Orientation
Assigning a Lead to a Group
Lead Groups specify which Leads should be programmed together in automated programming modes
(for example, to create an independent stimulation field).
Note: Multiple percutaneous Leads can be grouped together, but any paddle Lead must be in its own
Lead Group. In manual programming mode, you can program across Lead groups.