Configuring The Vercise Dbs System; Verifying The Vercise Dbs System Configuration - Boston Scientific Vercise DBS Mode D'emploi

Masquer les pouces Voir aussi pour Vercise DBS:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Vercise™ DBS Programming Manual

Configuring the Vercise DBS System

Verifying the Vercise DBS System Configuration

The Configuration (or Config) screen displays information about the system components. The
Stimulator and Remote Control product and serial numbers are automatically detected and
recorded. The Lead information is also displayed.
Note: The Configuration screen may only be accessed once a patient is selected.
To verify the Vercise DBS System configuration:
1
Click
If Lead information has not been entered, the information will be requested before the
Configuration screen is displayed.
2
Verify that the Lead information is correct.
3
Verify that the Stimulator model and serial number are correct.
4
Verify that the Remote Control model and serial number are correct.
5
Verify that the electrode labels correctly correspond to the left and right Leads (in the image
of the Leads to the right of the screen).
By default, Bionic Navigator assigns contacts 1-8 to the left Lead and contacts 9-16 to the
right Lead. If the contacts to not correspond correctly, see Swapping the Leads on page 18.
Vercise™ DBS Programming Manual
91053785-07
16 of 323
in the upper right-hand corner of any screen.

Publicité

Table des Matières
loading

Table des Matières