Change from combined ultrasound and
pneumatic lithotripsy to ultrasound mode only
Disconnect the Pneumatic Handpiece from the control unit and attach
the sterilization cap to the end of the hose.
Disconnect the Pneumatic Probe from the Pneumatic Handpiece
by applying slight axial pressure and turning the handpiece ¼ turn
counter clockwise.
Remove the angled suction connector and Pneumatic Handpiece from
the Ultrasound Handpiece.
20
FB-389_US_ed_2009-12 User man US handpiece BSC.indd 20
Attach the straight connector to the Ultrasound Handpiece and attach
the Stone Catcher device tubing to the straight connector.
It is necessary to change the angled suction connector with the
straight suction connector.
Always use the straight suction connector when working in
ultrasound mode only.