Bionic Navigator™ 3D 4.0 Programming Manual
Configuring the CP and Stimulator
The Configuration screen allows you to set up the CP and Stimulator for programming. On the
Configuration screen you can:
• Establish a programming connection between the CP and the Stimulator.
• Check the status of the programming connection.
• Identify Leads used and record where they were implanted.
• Set up the connection between the Leads and the Stimulator ports.
• Run LeadSync™ feature to identify and apply the relative offset of Leads.
• Check the impedances of each contact.
Placing the Stimulator into Pairing Mode (CP Mode)
1. Use the patient's Remote Control to enter the Stimulator into Pairing Mode (CP Mode).
a.
From the Lock screen, press and hold the
"Stimulator Ready for CP Connection" (Figure 3 and Figure 4).
Figure 3: Programs Button on Remote Control Figure 4: Stimulator Ready for CP Connection
b.
Alternatively, navigate to the
CP Mode: Stimulator (Figure 5).
select
Bionic Navigator™ 3D 4.0 Programming Manual
92395591-02 6 of 1595
button until the screen displays
Clinician Menu on the patient Remote Control and
Figure 5: CP Mode: Stimulator