Non-Critical Error Messages

Table 14: Non-Critical Error Message Table
Code
Error Title
Error Cause Text
200
Faulty Delivery
Unable to Read from Delivery Device
Device
Memory
205
Faulty Delivery
Unable to Write to Delivery Device
Device
Memory
210
Faulty Delivery
Faulty Delivery Device
Device
Thermocouple
211
Faulty Delivery
Faulty Delivery Device Trigger
Device
Signals
215
Faulty Delivery
Invalid Therapy Code
Device
218
Faulty Delivery
Delivery Device Impedance Error
Device
219
Faulty Delivery
Delivery Device Frequency Error
Device
220
Expired Delivery
Maximum Full Treatments Exceeded
Device
225
Faulty Delivery
Delivery Device is Permanently
Device
Disabled
230
Expired Delivery
Maximum Vapor Time Exceeded
Device
235
Prime Failed
Low Temperature (Prime)
236
Pre-Treatment Failed
Low Temperature (Pre-Treatment)
240
Prime Failed
Low Water Pressure (Prime)
241
Prime Failed
High Water Pressure (Prime)
242
Pre-Treatment Failed
Low Water Pressure (Pre-Treatment) Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime
243
Pre-Treatment Failed
High Water Pressure (Pre-
Treatment)
245
Prime Failed
High Temperature (Prime)
246
Pre-Treatment Failed
High Temperature (Pre-Treatment)
255
Treatment Halted
Low Temperature (Treatment)
260
Treatment Halted
High Water Pressure (Treatment)
265
Treatment Halted
Low Water Pressure (Treatment)
270
Syringe is Empty
Syringe Empty
275
Prime Failed
Syringe Water Fill Error
280
Treatment Halted
Elevated Coil Temperature
290
Faulty Delivery
High Temperature (Idling)
Device
291
Faulty Delivery
High Water Pressure (Idling)
Device
295
Faulty Delivery
Needle Deployment Error
Device
296
Faulty Delivery
Needle Retraction Error
Device
300
Saline Pump Error
Saline Pump Encoder Error
325
Confirm Bladder
Saline Instilled Limit Exceeded
Drain
Error Message
Remove and reconnect Delivery Device electrical cable. If problem persists, replace Delivery Device.
Remove and reconnect Delivery Device electrical cable. If problem persists, replace Delivery Device.
Replace Delivery Device.
Replace Delivery Device.
Remove and reconnect Delivery Device electrical cable. If problem persists, replace Delivery Device.
Replace Delivery Device.
Replace Delivery Device.
Replace Delivery Device.
Replace Delivery Device.
Replace Delivery Device.
Replace Delivery Device.
Replace Delivery Device.
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime
Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace
Delivery Device.
Check water line for kinks. Resume priming. If problem persists, replace Delivery Device.
Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace
Delivery Device.
Check water line for kinks. Resume pre-treatment vapor cycle. If problem persists, replace
Delivery Device.
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime
Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace
Delivery Device.
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime
Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace
Delivery Device.
Retract the needle and remove the Delivery Device from patient. Replace Delivery Device.
Check water line for kinks. Resume treatment. If problem persists, replace Delivery Device.
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime
Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace
Delivery Device.
Retract the needle and remove the Delivery Device from patient. Replace syringe and reprime
Delivery Device.
Refill syringe and reprime Delivery Device.
Partial treatment delivered. Check syringe and water line for bubbles or leaks. If no bubbles or leaks
are found, resume treatment. If problem persists, replace Delivery Device. If bubbles are found, replace
syringe and reprime Delivery Device. If leaks are found, replace Delivery Device.
Replace Delivery Device.
Replace Delivery Device.
Ensure needle is retracted. Replace Delivery Device.
Reattempt needle retraction. If problem persists, retract needle manually and replace Delivery Device.
Ensure Delivery Device saline flush line is correctly inserted into saline pump and pump door is closed. If
problem persists, contact Technical Support.
Saline instilled limit exceeded. Please confirm the physician has drained the bladder.