After performing a pre-treatment vapor cycle, a screen will be displayed indicating the
generator is ready for treatment.
Figure 16: Ready for Treatment Screen
A treatment is initiated by activating flush, deploying the needle, and pulling in and
holding the vapor activation button. When a treatment is in progress, the time clock will
count in whole seconds to the maximum preset treatment time. After treatment time
has elapsed, the generator will automatically end the treatment and will once again be
ready for treatment. Releasing and reactivating the vapor activation button will initiate
another treatment after the required rest period has elapsed.
Figure 17: Treatment in Progress Screen
If at any time the generator is not ready to perform a treatment (e.g. rest period between
treatments is in effect, etc.) the screen specified Waiting as depicted below and is
grayed out.
Figure 18: Pending Ready Indicator on Screen
Black (K) ∆E ≤5.0
The generator will automatically be ready for treatment (Figure 16) when the pending
condition is resolved. Monitor the ALERTS message area during treatment and take
action when specified to do so.
Turbo Flush
If visualization becomes cloudy during the procedure, Turbo Flush can be activated to
improve the visualization by increasing the saline flow rate. To activate Turbo Flush,
double tap flush activation button and hold. Treatments will not be performed during
this mode.
When visualization has been cleared, turn off Turbo Flush by releasing the button.

Connecting New Delivery Device During Treatment Session

The generator can detect if a new delivery device has been connected to the generator
during a therapy session.
If a new Delivery Device is connected during the therapy session, select New to create
a new procedure record or Continue to continue with current procedure record.
Figure 19: New Delivery Device
Turning Off the Rezūm™ Generator
In the event of power failure, the generator will automatically shut off. Turn the power
button off. Please remove the Delivery Device from the patient immediately per instructions
in the DFU. Turn on again to restart the generator to begin a new therapy session.
Caution: Before moving this generator, turn the power OFF, remove all accessories
from the patient, and unplug the power cord from the outlet.
1. Turn off the generator by pushing the bottom Power button located on the front of
the generator.
O
Figure 20: Power Button
Note: In the event of loss of power, a new procedure record is created. Previously
completed treatments will be saved in the prior procedure record.
Rezūm Generator Options Menu Items
To configure the generator and setup before the therapy session begins, select the
Options button on the lower toolbar.
Figure 21: Options Menu Button
The following options (if applicable) are available and described in further detail in the
sections below: