Figure 43: Procedure Summary
Select devices from the list using the check boxes on the left of the screen, and then select
View. The generator will populate the summary report with the list of selected devices.
Figure 44: Populating List
Figure 45: Procedure Summary - View
Use the scroll bar to view the entire summary.

Treatment Monitor

This screen is password protected and accessed by Boston Scientific personal only.

Treatment Config

This screen is password protected and accessed by Boston Scientific personal only.

Servicing

This screen is password protected and accessed by Boston Scientific personal only.

PROPER CARE AND CLEANING

Overview

Proper care of the Rezūm™ Generator is very simple, yet it is an important factor in its
reliability. This section describes the proper care and cleaning required for the generator.
Warning: No modification of this equipment is allowed. Do not attempt to service or
maintain the generator while in use with a patient.
Caution: When any of the following occur, turn the power off, remove all accessories
from the patient, and unplug the power cord from the outlet.
• There is smoke or a strange odor leaking out of the generator.
• The generator has been dropped or impacted by an object.
• Liquid or foreign matter gets inside the generator.
• If you think the generator may have been damaged.
Black (K) ∆E ≤5.0

Recommended Care

Warning: Failure on the part of all responsible individuals, hospitals, or institutions,
employing the use of Rezūm Generator, to implement the recommended cleaning
schedule may cause equipment failure and possible health hazards. The manufacturer
does not, in any manner, assume the responsibility for performing the recommended
maintenance schedule. The sole responsibility rests with the individuals, hospitals, or
institutions utilizing the Rezūm Generator.
Warning: Do not submerge the device in liquids or pour cleaning liquids over, into or
onto the generator.
Warning: Before conducting proper care and cleaning, turn the power OFF and unplug
the power cord from the outlet to prevent electric shock.
Caution: To prevent damage to generator, do not clean any part of the device with
phenolic compounds. Do not use abrasive or flammable cleaning agents. Do not
steam, autoclave, or gas-sterilize the generator.
Caution: Using this generator with the ventilation ports blocked could cause a
breakdown. Clean this generator with care.
Caution: Only approved equipment and accessories shall be connected to the generator.
Caution: The Rezūm Generator USB port is only intended for use by authorized service
personnel or to export treatment data.
To ensure the Rezūm Generator is always functional when required, Boston Scientific
recommends performing the following routine activities:
• Performing a Visual Inspection.
• Cleaning the Rezūm Generator.
• Routine Care per checklist in this section.
It is important that the generator is stored at room temperature if it is expected to be used.
The Rezūm Generator requires no calibration.

Visual Inspection

Warning: After the visual inspection, if the device is damaged or a message indicating the
generator is not to be used, take the generator out of service and call Customer service.
The generator should be carefully inspected prior to installation and use.
• Carefully inspect the generator case for stress or physical damage.
• Inspect all external connections for loose connectors.
• Inspect the external power cord for damage or cracking.
• Inspect the display for marks, scratches, or other damage.
• Verify that the Product label on the device is clearly legible and present.
Cleaning the Rezūm Generator
It is recommended that the generator be inspected after each use according
to the Checklist in this manual and cleaned when appropriate. Listed below are
recommendations for cleaning the generator. The generator does not need to be
sterilized before or after use.
Warning: Do not submerge the device in liquids or pour cleaning liquids over, into or
onto the generator.
Caution: After cleaning, allow complete drying before plugging in to an outlet by
wiping with a dry, soft cloth.
Caution: Do not soak the generator or accessories in any medical liquid. Also, keep
liquids out of the generator and accessories.
Caution: When using disinfectant solutions, follow the manufacturer's directions.
Caution: Using this generator with the air vent blocked could cause a breakdown.
Clean this generator with care.
Caution: To prevent damage to equipment, do not clean any part of the generator
with phenolic compounds. Do not use abrasive or flammable cleaning agents. Do not
steam, autoclave, or gas-sterilize the generator.

Recommended Cleaning Products

The following cleaning products may be used to clean the exterior surfaces of the generator:
• Water
• 70% Isopropyl Alcohol
• Super Sani-Cloth® Germicidal Disposable Wipes by PDI only
14