Error Message Table

The following tables list all error messages that are displayed by the generator. Follow error message instructions to resolve the error.

Critical Error Messages

Table 13: Critical Error Message Table
Code
Error Title
Error Cause Text
400
Generator
RF Power Supply Error
Error
405
Generator
Power Supply
Error
Initialization Error
425
Generator
SBC Communications
Error
Critical Timeout
430
Generator
RF Power Supply
Error
Communication Error
435
Generator
MCU Processing Error
Error
440
Generator
CPLD Self-Test Error
Error
450
Generator
Delivery Device Interface
Error
Self-Test Error
455
Generator
Saline Pump Self-Test
Error
Error
460
Generator
Syringe Pump
Error
Self-Test Error
465
Generator
Water Pressure
Error
Self-Test Error
470
Generator
Delivery Device
Error
Temperature Excessive
475
Generator
Software Compatibility
Error
Self-Test Error
480
Generator
Sensor Interface Error
Error
485
Generator
Internal Generator
Error
Temperature Error
490
Generator
RF Power Supply
Error
Operational Error
495
Generator
RF Power Supply Self-
Error
Test Error
35000
Generator
GUI Program Files
Error
Corrupted
35001
Generator
Unexpected GUI
Error
Program Exit
35002
Generator
Unable to Start
Error
GUI Program
40000
Generator
GUI Unable to
Error
Communicate with MCU
41020
Generator
MCU Reboot Detected
Error
Black (K) ∆E ≤5.0
Error Message
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact
Technical Support.
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact
Technical Support.
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact
Technical Support.
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact
Technical Support.
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact
Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Retract the needle and remove the Delivery Device from patient. Power off/on the Generator. Replace Delivery Device.
Power off/on the Generator. If problem persists, contact Technical Support.
Retract the needle and remove the Delivery Device from patient. If problem persists, contact Technical Support.
Retract the needle and remove the Delivery Device from patient. Power off the Generator and allow it to cool down before
using again.
Retract the needle and remove the Delivery Device from patient. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Power off/on the Generator. If problem persists, contact Technical Support.
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact
Technical Support.